Franck-Hertz  :   Xenon

Image #2

 

content.gif (1503 bytes)
 Code Number :   7B30.12

Disclaimer:

Reprinted by permission of Dick Berg, University of Maryland, for use on this website.

The demonstrations contained and referenced herein are listed for the purposes of cataloging and describing physics demonstrations which should be conducted only under the direction of a trained instructional support professional or physicist. These demonstrations are not presented for the purpose of being conducted by persons unconnected to this Facility and/or persons not consulting with or being supervised by the recognized instructional support professional or physicist and his/her staff. The University is responsible only for those demonstrations carried out using its own equipment using established safety and scheduling policies, and bears no responsibility for those choosing to use this source material for their own purposes. All demonstrations described and contained herein are public domain, and can also be found in reference materials in libraries, bookstores, and electronic sources.

Further information regarding legal liability in use of demonstrations and labs will be found on the web site Injuries in School/College Laboratories in USA.

The University of Iowa Disclaimers:  University of Iowa Disclaimer All Rights Reserved..

Condition :   Excellent  
Principle :   Critical Potential in a Gas   
Area of Study :  Modern Physics  
Equipment :   Franck - Hertz Xenon Tube Unit, 0 to 20 VDC Power Supply, 2 - Variacs, 2 - 6 volt Transformers, Volt and Amp Meters, Oscilloscope.

  
Procedure :   This version of the Franck - Hertz experiment is built around the 2D21 electron tube that has an appreciable amount of xenon gas present.  Since we like to show our peak on an oscilloscope instead of plotting it out, our hookup is a little different than that described in the instruction booklet in the files. We impress a 60 cycle, three volt ripple on the 0 to 12 VDC side of the circuit. Run the DC power supply at about 4 to 5 volts. Usually the ripple can be held to around 2 volts when everything stabilizes. The meter amplifier switch must be to Experiment 1 on the Franck - Hertz unit. Some tweaking of the power supply voltages and the control knobs on the Franck - Hertz unit will probably be necessary to obtain the best trace. Only one peak will be obtained with this unit.  
conbot.gif (53 bytes)


   References
 

"Ionization Potential of Xenon", AJP, 33 (5), May 1965, p.xvii